Plasma Technology for Health

Suche
Close this search box.

To compare the tissue tolerance and efficacy of two wound antiseptics with tissue-tolerable plasma (TTP) on enucleated contaminated eyes from slaughtered pigs in order to draw consequences for the use of TTP on wounds. The corneas of extracted eyes were contaminated with Staphylococcus aureus or Pseudomonas aeruginosa. One and 10 min after application of 10% povidone (PVP)-iodine and 0.04% polyhexanide, respectively, the eyes were rinsed with inactivating solution. To test TTP, the plasma pen meandered over the eyes at a speed of 30 mm/s and a distance of 5 mm; the eyes were then rinsed with balanced salt solution. The reduction factor was calculated by the difference between the logarithm of colony-forming units in the rinse before and after antisepsis or TTP application. The efficacy of TTP (reduction factor 2.4-2.9) was significantly higher (p < 0.001) than that of PVP-iodine and polyhexanide (reduction factor 1.7-2.1). TTP is more effective than the tested wound antiseptics. The lack of histological damage to the eyes of slaughtered pigs would seem to make its use as a wound antiseptic a viable alternative. In contrast to antiseptics, it supplies additional energy in the form of heat, electric fields and radicals by TTP.